User Manual

Table Of Contents
Transfer Tab Patient Setup Functions
5 - 34 0070-10-0690-01 Panorama™ Operating Instructions
The destination tile
must have a device
assigned to perform
the transfer.
The destination tile does not
have a device assigned.
Select the OK button on the message
box and then select a destination tile
that has a device assigned.
Please touch tile to
which you wish to
transfer the patient.
A destination tile has not been
selected at the Destination
Panorama.
Select a destination tile that has a
device assigned.
Failed Transfer Out:
System Error.
(Source Panorama)
The selected Destination
Panorama device was not
properly defined in the
equipment list.
Repeat the transfer procedure and
ensure that the desired Destination
Panorama device is properly defined
in the Equipment List.
There is an interruption in the
network connection between
the Source and Destination
Panoramas.
Determine the cause of the
interruption and correct.
Failed Transfer In:
System Error.
(Destination Panorama)
The selected Destination
Panorama device was not
properly defined in the
equipment list.
Repeat the transfer procedure and
ensure that the desired Destination
Panorama device is properly defined
in the Equipment List.
There is an interruption in the
network connection between
the Source and Destination
Panoramas.
Determine the cause of the
interruption and rectify.
This Tab is in use by
another display. Can
not be updated from
more than one
display at the same
time.
An attempt was made to open
the Transfer tab on one
display when it is already
open on the other display.
Select the OK button and use the
open the Transfer tab.
At the destination tile, the
digital data is displayed
but the waveform is not
displayed
The destination tile was in
Standby mode when the
transfer completed.
Return to normal monitoring mode.
ID number truncated. More than 10 characters
were entered in the ID field at
the WMTS 608 bedside
device.
Do not enter more than 10
characters when using a WMTS 608
bedside device because only 10
characters will be transferred to the
central station.
The destination tile is blank
and all buttons on the
Transfer tab are
unavailable.
The destination tile was in a
Communications Lost
state when the transfer
completed.
Re-establish communication.
Waiting for Patient
message is displayed in
the heart rate digital tile.
The destination device has not
been activated.
Activate the destination device.
Available monitor list
is empty.
The Central To Central
Transfer Without Device
button was selected but there
are no other Panorama
devices on the network.
Try and transfer to another central
station on a diff
erent network. If
unsuccessful, contact the System
Administrator for assistance.
Central To Central Transfer (Continued)
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.