User Manual

Table Of Contents
Transfer Tab Patient Setup Functions
5 - 32 0070-10-0690-01 Panorama™ Operating Instructions
The destination tile is blank
and all buttons on the
Transfer tab are
unavailable.
The destination tile was in a
Communications Lost
state when the transfer
completed.
Re-establish communication.
Unable to perform
operation on tile:
Central to Central
Transfer Out in
progress
A Central to Central transfer
Out has been initiated at the
source central station but not
been accepted or rejected at
the destination central station
for the selected tile.
Accept or reject the transfer at the
destination central station, and then
initiate Room to Room Transfer
again.
A Central to Central transfer
Out is in process at the source
central station
Accept or reject the transfer at the
destination central station, and then
initiate Room to Room Transfer
again.
Unable to perform
operation on tile:
Central to Central
Transfer In in
progress.
A Central to Central transfer
In is in progress for the
selected tile, which is actively
monitoring a patient, and a
Room to Room transfer is
attempted into the same
selected tile.
Wait until the Central to Central
transfer In is complete, then attempt
transfer again.
A Central to Central transfer
In has been initiated at the
source central station but not
accepted or rejected at the
destination central station for
the selected tile, and another
Central to Central transfer is
attempted for another patient
into the same selected tile.
Wait until the Central to Central
transfer In is complete, then attempt
transfer again.
The Central to Central
Without Device button in the
Transfer Tab was selected
when a Central to Central
transfer In is in progress for
the selected tile, and the
selected tile is actively
monitoring a patient.
Wait until the Central to Central
transfer In is complete, then attempt
transfer again.
A Central to Central transfer
In is in progress for the
selected tile, which has an
assigned device that is not
actively monitoring a patient,
and a Room to Room transfer
for another patient is initiated
into the same selected tile.
Wait until the Central to Central
transfer In is complete, then attempt
transfer again.
The Waiting for
Patient message is
displayed in the heart rate
digital tile.
The destination device has not
been activated.
Activate the destination device.
Room To Room Transfer (Continued)
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.