Panorama™ Operating Instructions 0070-10-0690-01 5 - 27
Patient Setup Functions Transfer Tab
•The Reject button terminates the transfer and returns to the Transfer tab. The message
Transfer Out rejected by [the Destination Panorama name] Central. is
displayed in the system status line of the Source Panorama.
•The Cancel button returns to the Transfer tab without terminating the transfer.
NOTE: A Central To Central Transfer Without Device can fail if the
network connection is lost or if either Panorama loses
power during the process. If a Central To Central Transfer
Without Device fails while in progress, the patient
information from the source tile is saved to a Failed Transfer
List at the Source Panorama.
The Failed Transfer List stores a maximum of five failed
transfers. If the maximum is reached, the Central To Central
Without Device button is disabled. A Central To Central
Transfer Without Device cannot be initiated until at least one
of the failed transfers is either recovered, deleted or
transferred using one of the sidebar buttons in the Transfer
tab. For additional information, refer to ‘Sidebar Buttons’
on page 5-28.
5.2.3.3 Effects of the Transfer
The following is a list of the changes that occur at the destination tile when a Central To
Central Transfer Without Device is complete:
Display
The transferred patient's parameter display changes, but only when the destination
device does not support the source device parameters.
The transferred patient's parameter display waveform 1 changes to ECG II when the
destination device does not support the source device's waveform 1 parameter and has a
5-lead ECG cable.
The transferred patient's parameter display waveform 1 changes to the active lead when
the destination device does not support the source device's waveform 1 parameter and
has a 3-lead ECG cable.
The transferred patient's parameter display waveform 2 changes to OFF when the
destination device does not support the source device's waveform 2 parameter.
If the source device is a bedside monitor, the units for CO
2
and temperature will change
to that of the destination device.
If the destination device and the source device are different, the source devices sensor
display setup will be reset to the system default settings of the destination device.
The care group is set to the system default at the destination device.
Demographics
For all transfers to a bedside monitor, the bed ID and patient size change to those of the
destination device. The remaining demographic data only changes when the destination
device has experienced a Communications Lost event. When communication is
re-established, the remaining demographic data is taken from the destination device.