Pacer Filtering adjusts the time period during which pacemaker-induced ECG
complexes are filtered for ECG sampling. The preset options for the Pacer Filter are 0 to
80 ms (in increments of 20). The Pacer Filter is selected in the Pacer Setting dialog box
(shown in FIGURE 5-8).
FIGURE 5-8 Pacer Setting Dialog Box
To enter Pacer Settings:
NOTE: For additional information regarding lead placement
procedures for pacemaker patients, refer to “Monitoring a
Pacemaker Patient” on page 4-9.
1. From the Demographics tab, select the Pacer sidebar button. The Pacer Setting
dialog box is displayed.
2. Select the Pacer Enhancement button to enable or disable this setting for the patient.
A checkmark indicates that Pacer Enhancement is enabled.
An empty box indicates that Pacer Enhancement is disabled.
3. Select the Pacer Filter button until the desired setting is displayed. The selections for
Pacer Filter range from 0 to 80 ms (in increments of 20).
4. Accept or reject the selections made in the Pacer Setting dialog box.
•The Done button accepts the selections.
•The Cancel button discards the selections, closes the dialog box and returns to the
Demographics tab.
5.1.3.2 Arrhythmia Setup
Select the Arrhythmia sidebar button to:
Enable or disable arrhythmia analysis (for Telepack device only)
Adjust the system default V-Tach Rate, Asystole Delay, V-Tach Thresholds, and ECG
Noise Delay.
NOTE: The Arrhythmia sidebar button is not available at the
ViewStation or the WorkStation.