Installation Guide
Panorama WMTS Installation Guide 0070-00-0571 7 - 3
Appendix A
5. Verification of desired coverage and latency. Verification must be performed using an
ambulatory transceiver. Mark the status box in the table below for each successfully
completed test.
a. Main (A) Coverage - connect the A antenna leg to the Main input on the TIM
transceiver. Validate proper transmission throughout the A coverage area.
b. Diversity (B) Coverage - connect the B antenna leg to the Main input on the TIM
transceiver. Validate proper transmission throughout the B coverage area.
c. Full ECG Coverage Test - re-connect both antenna legs to their respective connectors
on the TIM transceiver. Establish an ECG and admit the transceiver. Verify the central
station receives an ECG waveform at all points in the covered area.
d. Latency Test - go to farthest antenna from the central station. Go to the extreme edge
of the reception area for this antenna. Initiate an arrhythmia and verify that the
Central alarms in less than 10 seconds. Document this time.
e. Transceiver Verification - ensure that all transceivers are functional.
LASER PRINTER
MODEL SERIAL NUMBER HW LAN ADDRESS CLAN IP ADDRESS
DEVICE TYPE SERIAL NUMBER DEVICE ID IP ADDRESS
• Transmit an ECG simulator from each ambulatory transceiver and verify that
the waveform is received at the assigned Central on the assigned channel.
• Transmit all appropriate waveforms from each instrument transceiver and
verify that all waveforms are received at the assigned Central on the assigned
channel.
ACTION STATUS COMMENTS
a.
Main A Test
b.
Diversity B Test
c.
ECG Coverage Test
d.
Latency Test (record
delay)
e.
Ambulatory
transceivers