Installation Guide
Configuring Bedside Monitors for WMTS telemetry Programming and Configuring Telemetry Devices
5 - 2 0070-00-0571 Panorama WMTS Installation Guide
5.1 Configuring Bedside Monitors for WMTS telemetry
Each bedside monitor being connected via telemetry must be configured to communicate to
the radio card.
Enter the Installation Menu on the bedside monitor.
1. Press and hold the Discharge key during power up.
2. Set WMTS Enabled to Yes
3. Save the information by pressing Save Current.
4. Turn off the bedside monitor.
Repeat the above procedure on all bedside monitors being connected via WMTS telemetry.
5.2 Assigning the Telemetry Server(s) to the Equipment
List
NOTE: Prior to programming the Telemetry Devices, the Wireless
Server(s) must be assigned to the each central station
equipment list.
1. Turn on the central station and Displays. Wait for the operating software to start.
2. Press the System Setup button.
3. Press the Installation Setup tab.
4. Enter the Password "System".
5. Press the Equipment Setup tab.
6. Press the New button.
7. Press Tile and select the tile to be assigned: None for VP Server, then press Done.
8. Press Type and select the device type to be entered: VP Server, then press Done.
9. Press Label and use the onscreen keyboard to type in a label for that tile. Press the
Enter key on the keyboard when the label is completed.
10. · Press Device ID and enter the Device ID of the VP Server being assigned to that tile.
11. · Press Done to complete the Server configuration
12. Repeat this procedure for all Wireless Servers to be assigned to the system.