Installation Guide
Panorama WMTS Installation Guide 0070-00-0571 2 - 3
System Installation Tools and Equipment Required
The Panorama Telemetry Server and TIM Transceiver must be located close enough in the
rack to be connected by the LVDS cable.
Using FIGURE 2-1 as a reference:
1. Connect one end of the cable to LVDS connector on the wireless server. The primary
server should be connected to DS1 on the rear of the wireless transceiver.
NOTE: Additional Servers will connect to the remaining connectors
DS2 through DS4 on the rear of the wireless transceiver
2. Connect the keyboard and mouse into their respective ports on the rear of the Wireless
Server.
3. Connect the Quad I/O cable and VGA cable into their respective ports on the rear of
the Server.
4. Terminate the unused ports on the Quad I/O cable with P/N 0992-00-0135
terminators.
5. Connect an AC Power cord to the Server.
6. Connect the AC cord to the UPS.
7. Connect the VGA cable to the display monitor. Connect the other end of the VGA cable
to monitor VGA input 1.
8. Connect an AC Power cord to the display monitor. Connect the AC cord to the UPS.
FIGURE 2-1
Local Installation (Server(s) Co-located with Central Station)
Direct connection of the Wireless (Telemetry) Server (P/N 0998-00-0206-01) to the central
station is via a crossover CAT5 patch cable (P/N 0012-00-1392-XX).
Keyboard
Connector
Mouse
Connector
Patient
Network
Te l em e try
Network
AC Power
Input
VGA Monitor
(Input #1)
Quad I/O
Connector
LVDS
Connector