User's Manual

Transfer Tab Patient Setup Functions
5 - 16 0070-00-0575-50 Panorama™ Operating Instructions
Demographics
For all transfers to a bedside monitor, the bed ID and patient size change to those of the
destination device. The remaining demographic data only changes when the destination
device has experienced a Communication Lost event. When communication is re-
established, the remaining demographic data is taken from the destination device.
For all transfers to a Panorama Telepack, the bed ID is cleared and the patient size
changes to Adult (the Telepack default). The remaining demographic data is preserved.
Alarms
The transferred patient's Alarm Limits, Alarm Delay and ST Alarm Delay change to those
of the destination device when Alarm Tracking mode is ON.
All alarms for the transferred patient become enabled, discontinuing all Alarm Suspended
settings.
5.2.3 Central To Central Transfer Without Device
A Central To Central Transfer Without Device is used to transfer patient data stored
at one Panorama Central Station to a different Panorama Central Station. Historical patient
data is preserved upon successful completion of the transfer.
The Central To Central Without Device button is enabled under the following
conditions:
When a patient tile actively monitoring a patient is selected as the source
When a patient tile actively monitoring a patient, but has been placed in Standby
mode, is selected as the source
When a patient tile actively monitoring a patient, but experiencing a Communication
Lost event, is selected as the source
NOTE: For information on changes that occur at the destination tile
when the transfer is complete, refer to ‘‘Effects of the
Transfer’’ on page 5-19.
5.2.3.1 Source Panorama
1. A prerequisite for the transfer is that certain patient demographic information must exist.
In the patient tile to be transferred, ensure that the appropriate information has been
entered into at least two of the three demographic fields (Last Name, First Name and ID)
on the Demographics tab.
2. Ensure that the device name for the Source Panorama is properly entered into the
equipment list at the Destination Panorama as described in the ‘‘Equipment Setup Tab’’
on page 9-21.
3. Ensure that the device name for the Destination Panorama is properly entered into the
equipment list at the Source Panorama as described in the ‘‘Equipment Setup Tab’’ on
page 9-21.
4. Select the patient tile to be transferred. Choose a patient tile by selecting its digital data
or waveform area.