User's Manual

Panoramaâ„¢ Operating Instructions 0070-00-0575-50 5 - 13
Patient Setup Functions Demographics Tab
5.1.4 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Demographics tab.
MESSAGE/ISSUE * REASON SOLUTION
Can not admit Patient to
this tile. No Device
attached to this tile
There is no device assigned to
the selected tile.
Select the OK button to close
the dialog box. Select a patient
tile that has a device attached
to it or attach a device to the
selected patient tile.
You have reached the
maximum number of
arrhythmia licenses
available
There are no more Arrhythmia
licenses available.
Select the OK button to close
the dialog box. Arrhythmia
Analysis cannot be selected for
a patient until it is discontinued
for another patient. The System
Administrator can purchase
additional Arrhythmia licenses.
The Pacer, Care Group,
and the Retrieve
Discharged Pt. sidebar
buttons are disabled.
The selected patient tile is in
Standby mode.
Remove the selected patient tile
from Standby mode.
The filter mode cannot be
Monitor when ST is on
Monitor mode cannot be
selected while ST is enabled.
Select the OK button to close
the dialog box. Turn off ST
select the correct filter mode.
Turning on ST will change
your ECG filtering to ST
mode
ST mode was selected while in
Monitor mode.
Select the OK button to close
the dialog box. Select the
correct filter mode.
You have reached the
maximum number of ST
licenses available
There are no more ST licenses
available
Select the OK button to close
the dialog box. ST Analysis
cannot be selected until it is
discontinued on another
patient. The System
Administrator can purchase
additional ST licenses.
Retrieve Discharged
Patient Failed: The
destination tile is
currently occupied
A patient cannot be retrieved
into an occupied tile.
Select the OK button to close
the dialog box. Select a vacant
patient tile before attempting to
retrieve a patient.
* Messages are shown in all bold text.