Demographics Tab Patient Setup Functions
5 - 10 0070-00-0575-50 Panorama™ Operating Instructions
2. Select a Care Group button.
3. Accept or reject the Care Group selection.
•The Done button accepts the Care Group selection.
•The Cancel button discards the selection, closes the dialog box and returns to the
Demographics tab.
For additional information regarding Care Groups, refer to the ‘‘Care Group Tab’’ on
page 9-2.
Settings (Telepack)
Select the Settings sidebar button to establish the type of ECG monitoring performed by the
Telepack.
•The Monitor mode establishes the frequency response range for non-ST ECG
monitoring.
•The ST mode establishes the frequency response range for ST diagnostic monitoring.
To enter settings for a Telepack:
1. From the Demographics tab, select the Settings sidebar button. The Settings dialog
box (FIGURE 5-8) is displayed.
FIGURE 5-8 Settings Dialog Box
2. Select the ST button to enable or disable ST analysis for the selected patient.
A checkmark in the ST box indicates that ST analysis is enabled.
An empty box indicates that ST analysis is disabled.
NOTE: If ST licenses are available, an ST license is automatically
assigned to the patient when ST is enabled
.