User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 5 - 5
Patient Setup Functions Demographics Tab
FIGURE 5-3 Height Dialog Box
2. Use the dial selector or the arrow buttons to select a value in the Height dialog.
3. Accept or reject the value in the Height dialog box.
•The Done button accepts the selection.
•The Cancel button discards the selection, closes the dialog box and returns to the
Demographics tab.
Weight Fields
The Weight fields display the patient weight in metric and standard units. The field on the
left is used to enter the patient’s weight. The unit of measure for this field is dependent on the
setting at the bedside monitor. The field on the right automatically displays the conversion of
the measurement. A patient’s weight can be entered at the bedside monitor or the Panorama
Central Station because the system enables an automatic bi-directional flow of information.
UNITS OF MEASURE INCHES (in) CENTIMETERS (cm)
Range 8.0 to 120.0 in 20.0 to 305.0 cm
Single arrow Step Setting 0.5 (8.0 to 20.0 in)
1.0 (20.0 to 120.0 in)
1.0 cm
Double arrow Step Setting 1.0 (8.0 to 20.0 in)
10.0 (20.0 to 120.0 in)
10.0 cm
Dial Selector
Field Name
Edit Box
Arrow Buttons