User's Manual

Lead Placement Lead Placement
4 - 12 0070-00-0575-50 Panorama™ Operating Instructions
IEC Modified Chest Lead (MCL) Monitoring with a 3-wire Lead Set
FIGURE 4-10 MCL Monitoring with a 3-wire Lead Set (IEC)
Some patients may require monitoring in a MCL configuration.
• Place R (red) electrode under patient left clavicle, mid-clavicular line within the rib cage
frame.
• Place L (yellow) electrode right sternal border, fourth intercostal space within the rib cage
frame.
• Place F (green) electrode on the patient’s lower left abdomen within the rib cage frame.
• Select ECG Lead I for MCL
1
monitoring. Lead I is the direct electrical line between the
R (red) electrode and the L (yellow) electrode.
• Select ECG Lead II for MCL
6
monitoring. Lead II is the direct electrical line between the
L (red) electrode and the F (green) electrode.