Panorama™ Operating Instructions 0070-00-0575-50 4 - 3
Lead Placement Electrode Patch Location
2. Attach the electrode patch to the skin at the prepared site. Smooth the electrode patch
down in a circular motion to ensure proper skin contact. If using soft gel electrodes,
never push down directly over the contact gel medium as this may displace the gel and
cause monitoring artifact. If using hard gel electrodes, it is recommended that during
application, the center of the electrode should be slightly pressed onto the skin to ensure
direct contact. Consult the electrode patch manufacturer’s instructions for specific use.
3. Secure the lead wires to the patient according to hospital practice. For additional
information see “Lead Placement” on page 4-4. If using a Panorama Telepack, always
secure the device to the patient according to hospital standard.
WARNING: Ensure that the ECG lead wires are neatly secured in such a
way to prevent the lead wires from encircling the patient’s
neck and causing possible strangulation.
NOTE: It is recommended that electrode patches be changed at
least every 24-36 hours to maintain proper contact with the
skin. Some patients may require electrodes to be changed
more often. Electrode patches are disposable and should be
applied only once. Try to avoid reusing the exact same
electrode site during reapplication. If an electrode becomes
wet with fluid, change the electrode patch.