Panorama™ Operating Instructions 0070-00-0575-50 2 - 19
Getting Started Quick Reference Guide
Muting Alarms
To mute a non-latched alarm (for example, IBP):
Select the Mute button in the alarming patient's tile to silence the alarm violation for the
configured time period.
For additional information on muting an alarm, refer to "VIEW/MUTE Button’’ on page 2-5.
To mute a latched alarm (for example, lethal arrhythmias):
Select the digital data area or waveform area of the alarming patient tile to acknowledge
the alarm AFTER the alarm condition is resolved.
NOTE: The audio and visual indicators for a latched alarm will
continue until the alarm is acknowledged, even if the alarm
condition no longer exists.
2.4.6 Viewing List Trends
The List Trends display presents all available digital data for a patient, in a list format.
To view List Trends at the Panorama Central Station:
1. Select the VIEW button in the desired patient tile.
2. Select the Trends tab.
Select the Print More button or the Print Current button to print the list trends.
To adjust the Trend Storage period:
1. Select the Patient Setup menu button.
2. Select the Display tab.
3. Select Trend Storage sidebar button.
For additional information on Viewing List Trends, refer to "List View of the Trends Tab’’ on
page 7-8.
2.4.7 Viewing Events
The Events tab displays all events and related waveform data (maximum of 1,000 events)
for a patient.
To view Events at Panorama Central Station:
1. Select the VIEW button in the desired patient tile.
2. Select the Events tab.
Select the Print More button or the Print Current button to print the events list.
3. Select the Event Filter sidebar button to adjust the filter of the event list.
For additional information on events, refer to ‘‘Events Tab’’ on page 7- 16.