User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 15 - 29
Specifications Telepack Specifications
TABLE 15-5
GUIDANCE AND DATASCOPE CORP. DECLARATION - ELECTROMAGNETIC EMISSIONS
It is important that the Telepack is used in the electromagnetic environment specified in the following
table.
EMISSIONS
TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions
CISPR 11
Group 1 The Telepack uses RF energy only for its internal function
and external data transmission. However, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The Telepack is suitable for use in all establishments
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes
Harmonic emissions IEC 61000-3-2 N/A
IEC 61000-3-3 Voltage fluctuations/
Flicker emissions
N/A
TABLE 15-6
GUIDANCE AND DATASCOPE CORP. DECLARATION - ELECTROMAGNETIC IMMUNITY
It is important that the Telepack is used in the electromagnetic environment specified in the following
table.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines ±1
kV for input/
output lines
Not applicable * Mains power quality should be typical
of a commercial or hospital
environment.
Surge IEC
61000-4-5
±1 kV
differential
mode ±2 kV
common mode
Not applicable * Mains power quality should be typical
of a commercial or hospital
environment.
U
T
is the a.c. mains voltage prior to application of the test level.
* Telepack operates only from the internal bat