Specifications Telepack Specifications
15.11 Telepack Specifications
The following are the specifications for the Telepack:
15.11.1 Safety Designations
Safety designations as per IEC 601-1 Standard:
15.11.2 Panorama Server Analysis
The Telepack device collects ECG information and transfers it to the Panorama Telemetry
Server for analysis. In contrast, the Passport 2
®
collects and analyses patient information and
then transfers its analysis to the Panorama Central Station; therefore, no calculations are
performed by the Panorama Server for the Passport 2. The specifications of the Panorama
Telemetry Server relative to the Telepack are described in the following table.
TABLE 15-1
SPECIFICATION DESCRIPTION
Pacer Detection Provides pacer detection capability
Acquired Leads ECG 3-Lead: I, II, III (one vector at a time)
ECG 5-Lead: Three vectors (I, II, III, aVR, aVF, V(n))
Battery Type and Runtime
(Telepack-2.4)
24 hrs minimum using (2) AA alkaline Cells with ECG at a HR of
60 bpm common to the Panorama Central Station
Battery Type and Runtime
(Telepack-608)
40 hrs minimum using (2) AA alkaline Cells with ECG at a HR of
60 bpm common to the Panorama Central Station
46 hrs minimum using (2) AA Lithium Cells with ECG at a HR of
60 bpm common to the Panorama Central Station
Battery Shelf Life The battery shelf life is indicated on the battery casing.
Lead Current Max lead current: 25nA
TABLE 15-2
DESIGNATION DESCRIPTION
Degree of protection against
electric shock
ECG-Type CF defibrillation protected.
Supply Connection 3 VDC Internal Battery
Mode of Operation Continuous
Protection Against Hazards of
Explosion
Not Protected (Ordinary)
Protection Against Ingress of
Liquids
IPx7
Degree of Electrical Connection
between Equipment and Patient
Equipment designed for direct electrical connection to the patient
Degree of Mobility Transportable, Intra Hospital
TABLE 15-3
SPECIFICATION DESCRIPTION
Performance with Telepack Supports a maximum of 16 devices, 8 of which can be Telepack
Displayable Leads 3-Lead - 1, II, III (one vector at a time) (for Telepack-608 only)
5-Lead - I, II, III, aVR, aVL, aVF, V1
Notch Filter Set to 60 Hz