User's Manual

Panorama Wireless Transceiver/Splitters/Antennas Specifications
15 - 24 0070-00-0575-50 Panorama™ Operating Instructions
15.10.4.15 Safety
Enclosure Risk Current
3.5 mA max. per EN 60950
Dielectric Withstand
The Panorama Wireless Transceiver and its components shall withstand and operate as
specified after application of 1500 V RMS at 50 or 60 Hz for 1 minute from AC mains
hot or neutral to chassis.
Ground Resistance
For items that have a connection to AC mains power:
The Panorama Wireless Transceiver and its components shall have a ground resistance of
0.1 ohm from the AC mains power inlet module’s ground contact pin to any exposed metal
part, which may become energized when, measured per IEC 950. A ground resistance of up
to 0.2 ohm is allowed when measured from the U blade of the supplied AC line cord to any
exposed metal part, which may become energized.
15.10.4.16 Radio Certification
Panorama Wireless Transceiver
FCC ID: DXXVT0190-00000