User's Manual

Panorama Wireless Transceiver/Splitters/Antennas Specifications
15 - 22 0070-00-0575-50 Panorama™ Operating Instructions
CAN/CSA Standard C22.2 No. 950-95, Third Edition, and UL 1950, which does not
require the shock and vibration testing.
15.10.4.9 Fluid Spill Resistance
UL 1950, which does not require fluid spill resistance testing.
15.10.4.10 Surface Temperature
Temperature does not exceed the values shown in table 16, parts 1 and 2 of UL 1950:
1995.
15.10.4.11 Tip Over
UL 1950: 1995, section 4.1.1: Unit does not overbalance when tilted to an angle of 10°
from normal upright position.
15.10.4.12 Rigidity and Strength of Enclosure
UL 1950: 1995 4.1.2, 4.2.2, 4.2.3, 4.2.4, and 4.2.5
15.10.4.13 External Connector Compatibility
Clause 4.3.17 of the CAN/CSA Standard C22 no 950-95, third edition and UL 1950
third edition, standard: “Safety of Information Technology Equipment.”
15.10.4.14 Electromagnetic Compatibility
The electromagnetic compatibility (EMC) requirements are summarized as follows:
Radiated Emissions
EN55022:1998 Class A Radiated
(RF) Conducted Emissions
EN55022:1998 Class A Conducted
Radiated Susceptibility
EN 61000-4-3:1998: 3 V/M: 26 MHz – 1 GHz 80% AM
Conducted Susceptibility
EN 61000-4-6:1996
0.15-80 MHz
3V
RMS
, 80% AM
Electro-Static Discharge (ESD)
EN 61000-4-2; 2, 4, and 6 contact and 2,4,6 and 8 kV air