User's Manual

Panorama Wireless Transceiver/Splitters/Antennas Specifications
15 - 22 0070-00-0575-50 Panorama™ Operating Instructions
• CAN/CSA Standard C22.2 No. 950-95, Third Edition, and UL 1950, which does not
require the shock and vibration testing.
15.10.4.9 Fluid Spill Resistance
• UL 1950, which does not require fluid spill resistance testing.
15.10.4.10 Surface Temperature
• Temperature does not exceed the values shown in table 16, parts 1 and 2 of UL 1950:
1995.
15.10.4.11 Tip Over
• UL 1950: 1995, section 4.1.1: Unit does not overbalance when tilted to an angle of 10°
from normal upright position.
15.10.4.12 Rigidity and Strength of Enclosure
• UL 1950: 1995 4.1.2, 4.2.2, 4.2.3, 4.2.4, and 4.2.5
15.10.4.13 External Connector Compatibility
• Clause 4.3.17 of the CAN/CSA Standard C22 no 950-95, third edition and UL 1950
third edition, standard: “Safety of Information Technology Equipment.”
15.10.4.14 Electromagnetic Compatibility
• The electromagnetic compatibility (EMC) requirements are summarized as follows:
Radiated Emissions
• EN55022:1998 Class A Radiated
(RF) Conducted Emissions
• EN55022:1998 Class A Conducted
Radiated Susceptibility
• EN 61000-4-3:1998: 3 V/M: 26 MHz – 1 GHz 80% AM
Conducted Susceptibility
• EN 61000-4-6:1996
0.15-80 MHz
3V
RMS
, 80% AM
Electro-Static Discharge (ESD)
• EN 61000-4-2; 2, 4, and 6 contact and 2,4,6 and 8 kV air