User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 15 - 13
Specifications Panorama Telemetry Server
15.4.7 Storage Temperature
-20 to +60 º C
15.4.8 Storage Humidity
5% to 80% Relative Humidity Maximum, non-condensing, maximum wet bulb 35 º C
15.4.9 Storage Altitude
1060 to 700 hPa (-1250 to 9889 feet ASL) (-380 - 3014 m) (795 to 525 mm Hg)
15.4.10 Surface Temperature
Temperature does not exceed the values shown in table 16, parts 1 and 2 of EN 60950:
1992.
15.4.11 Tip Over
EN 60950:1992, section 4.1.1: Unit does not overbalance when tilted to an angle of 10°
from normal upright position.
15.4.12 Rigidity and Strength of Enclosure
External enclosures to meet steady force of 250 N ± 10 N for a period of 5 s. Parts of
enclosure located in operator access area to meet steady force of 30 N ± 3 N for a period
of 5 s.
15.4.13 External Connector Compatibility
Clause 4.3.17 of the CAN/CSA standard C22 no 950-95, third edition and UL 1950 third
edition.
15.4.14 Alarms: Audio and Visual
EN 475:1995
15.4.15 Electromagnetic Compatibility
The electromagnetic compatibility (EMC) EN 60601-1-2:2001 requirements are summarized
as follows: