User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 15 - 9
Specifications Panorama Central Station
Radiated Emissions
EN55022:1998 Class A Radiated
(RF) Conducted Emissions
EN55022:1998 Class A Conducted
Radiated Susceptibility
EN 61000-4-3:1998: 3 V/M: 26 MHz – 1 GHz 80% AM
Conducted Susceptibility
EN 61000-4-6:1996
0.15-80 MHz
3V, 80% AM
Electro-Static Discharge (ESD)
EN 55024:1998
EN 61000-4-2; 2 and 4 contact and 2,4,6 and 8 kV air
Electrical Fast Transient (EFT/B)
EN 55024:1998
EN 61000-4-4:1995 2kV supply and control 1kV data and signal
Surge Immunity
EN55024:1998
EN 61000-4-5:1995 2kV common mode; 1kV differential to AC line
Magnetic Emissions
MIL-STD-461D
RES101 30 Hz to 100KHz @ 7cm
Magnetic Immunity
EN55024:1998
EN 61000-4-8: 1994 50 HZ, 1 A/m
Voltage Dips, Short Interruptions and Voltage Variations
EN 55024:1998
EN 61000-4-11:1994
Harmonics