User's Manual

ECG Derived Heart Rate Meter Performance Requirements Specifications
15 - 6 0070-00-0575-50 Panorama™ Operating Instructions
15.2.6 IBP Heart Rate Meter
15.2.7 Temperature Performance Requirements
15.2.8 ECG Respiration Performance Requirements
15.2.9 SpO
2
15.2.10 CO
2
Performance Requirements
15.2.11 CO
2
Alarm
Range: 30 to 300 bpm Adult/Pediatric
30 to 333 bpm Neonatal
Resolution: 1 bpm
Scale: Selectable Celsius or Fahrenheit
Range: 15 °C to 45 °C
59 °F to 113 °F
Resolution: 0.1 °C
0.1 °F
Range: 4 to 199 breaths per minute
Resp. Scale: 1, 2, 3, 4 or 5 with standard ECG cable
SpO
2
Range: 70% to 100%
Pulse Rate Range: 30 to 235
Range: 0 - 13%s
Respiration Rate Range: 0 - 150 breaths/minute
ET CO
2
High Alarm Range: 2 - 10%
ET CO
2
Low Alarm Range: 1 - 6%
Insp. CO
2
High Alarm Range: 1 - 4%