User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 15 - 5
Specifications ECG Derived Heart Rate Meter Performance Requirements
User Selectable ST
15.2.3 Arrhythmia Analysis
Arrhythmia analysis is disabled in neonatal mode. Arrhythmia analysis will identify
ventricular arrhythmia.
The following arrhythmia calls will be displayed:
15.2.4 NIBP Performance Characteristics
Systolic Pressure Measurement
15.2.4.1 Diastolic Pressure Measurement
15.2.4.2 Pulse Rate
15.2.5 IBP Pressure Range
Measurement Points: 40, 60 and 80 ms after J point (heart rate independent)
Or 60/80 Heart Rate Dependent
Asystole Triplets PVCs per minute
Missed Beat Runs Ventricular Tachycardia
Irregular Heart Rate Bigeminy Ventricular Fibrillation
Couplets Trigeminy
Range: 55 to 235 mmHg in Adult mode
55 to 160 mmHg in Pediatric mode
45 to 120 mmHg in Neonatal mode
Resolution: 1 mmHg
Range: 30 to 200 mmHg in Adult mode
30 to 150 mmHg in Pediatric mode
20 to 100 mmHg in Neonatal mode
Resolution: 1 mmHg
Range: 35-245 bpm, for Adults/Pediatric
70– 245 bpm, for Neonate
Resolution: 1 bpm
Range: 30 to 300 mmHg after zeroing at the Passport