User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 15 - 3
Specifications ECG Performance Requirements
Waveform Displays Each waveform in the All Strips, Event List Zoom-In, Full
Disclosure Zoom-In and ST reports will be displayed on
a calibrated grid (5mm x 5mm) that conforms to ANSI/
AAMI EC13: 2002.4.2.9.7d.
Each second of the waveforms in the All Strips, Event List
Zoom-In, Full Disclosure Zoom-In and ST reports will be
marked with a time annotation marker that conforms to
ANSI/AAMI EC13: 2002.4.2.9.7e.
Heart Rate Meter Accuracy and
Response to Irregular Rhythm
Per ANSI/AAMI EC13-2002 Section 4.1.2.1e.
paragraph 3a, Ventricular Bigeminy, the Heart Rate
meter response is in the range of 39 bpm to 80 bpm.
Per ANSI/AAMI EC13-2002 Section 4.1.2.1e.
paragraph 3b, Slow alternating Ventricular Bigeminy,
the Heart Rate meter response is in the range of 45 bpm
to 62 bpm.
Per ANSI/AAMI EC13-2002 Section 4.1.2.1e.
paragraph 3c, Rapid alternating Ventricular Bigeminy,
the Heart Rate meter response is constant at 60 bpm.
Per ANSI/AAMI EC13-2002 Section 4.1.2.1e.
paragraph 3d, Bidirectional systoles, the Heart Rate
meter response is constant at 90 bpm.
Time to Alarm for Tachycardia The alarm response time for tachycardia is less than 5
seconds in accordance with ANSI/AAMI EC13-2002
Section 4.1.2.1g.
Pacer Rejection The heart rate meter rejects pulses of amplitude signals
from ±2mV to ±700mV and duration 0.1ms to 2ms with
no tail according to ANSI/AAMI EC13-2002 Section
4.1.4.1 (No tails).
Pacemaker Pulse Rejection
Capability
The pacer detector’s slew rate is 6.024V/s
NOTE: The heart rate meter does not reject Pacemaker pulses with
tails.
NOTE: The heart rate meter does not reject Pacemaker with fast
ECG signals.
NOTE: There is no Pacemaker pulse auxiliary output.