User's Manual

ECG Performance Requirements Specifications
15 - 2 0070-00-0575-50 Panorama™ Operating Instructions
15.1 ECG Performance Requirements
Three Lead Displayable Leads: I, II, III (one vector at a time)
Five Lead Displayable Leads: I, II, III, aVR, aVL, aVF, V1
Twelve Lead Displayable Leads: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5 and V6
Standardizing Voltage: The ECG size is indicated by a scale bar displayed on
the left of the waveform window with a scale indicating
the height of the bar in mV. Printed output to have
standardizing pulse printed at beginning of strip
ANSI/AAMI EC13-2002, 4.9.9
Time base selection
(Non-permanent display)
25 mm/s
ANSI/AAMI EC13-2002, 4.9.6 (b)
Time base error
(Non-permanent display)
±10%
ANSI/AAMI EC13-2002, 4.9.6 (b)
Impulse Response 0.1 mV max
0.3 mV/sec max slope
ANSI/AAMI EC13-2002, 4.2.9.8 (c)
Aspect Ratio 0.4 ± 0.08 sec/mV (met by 25 mm/s)
ANSI/AAMI EC13-2002, 4.2.9.7 (f)
Overall System Error Greater of 5% or 4 cuv
ANSI/AAMI EC13-2002-4.2.9.8 a
ECG Input dynamic range +/- 5mV max
ANSI/AAMI EC13-2002-4.2.9.1
Channel Width 30 mm. min
ANSI/AAMI EC13-2002-4.2.9.7 a