User's Manual

Technical Alarms and Events Alarms and Events
14 - 14 0070-00-0575-50 Panoramaâ„¢ Operating Instructions
SPO
2
Unrecognized
Sensor
The SPO
2
Unrecognized Sensor event will be posted in the
Events database if an unrecognized SPO
2
Masimo sensor is detected
by the system.
The SPO
2
Unrecognized Sensor message will be shown in the Patient
Status line in white text.
SPO
2
Sensor Off The SPO
2
Sensor Off event will be posted in the Events database if
the SPO
2
Masimo sensor is off.
The SPO
2
Sensor Off message will be shown in the Patient Status line
in white text.
SPO
2
No Sensor The SPO
2
No Sensor event will be posted in the Events database if
the SPO
2
Masimo sensor is not plugged in.
The SPO
2
No Sensor message will be shown in the Patient Status line
in white text.
SPO
2
Sensor Failed The SPO
2
Sensor Failed event will be posted in the Events
database if communication with the SPO
2
Masimo sensor failed.
The SPO
2
Sensor Failed message will be shown in the Patient Status
line in white text.
SPO
2
Interference The SPO
2
Interference event will be posted in the Events database
if there is interference in the communication with the SPO
2
Masimo
sensor.
The SPO
2
Interference message will be shown in the Patient Status
line in white text.
SPO
2
(NELLCOR) TECHNICAL EVENTS
SPO
2
Motion The SPO
2
Motion event will be posted in the Events database if the
SPO
2
Nellcor sensor detects motion.
The SPO
2
Motion message will be shown in the Patient Status line in
white text.
SPO
2
Check Sensor The SPO
2
Check Sensor event will be posted in the Events
database if a problem is detected with the SPO
2
Nellcor sensor.
The SPO
2
Check Sensor will be shown in the Patient Status line in
white text.
SPO
2
No Pulse The SPO
2
No Pulse event will be posted in the Events database if a
a pulse is not detected by the SPO
2
Nellcor sensor.
The SPO
2
Check Sensor message will be shown in the Patient Status
line in white text.
IBPX TECHNICAL EVENTS*
IBPx (Labels include: Art,
UI, LV, PA, CVP, ICP, LA,RA)
Not Calibration/Zeroed
The IBPx Not Calibration/Zeroed event will be posted in the
Events database if the IBP device is not zeroed.
There can be up to four IBP measurements. The IBPx Not Calibration/
Zeroed message(s) will be shown in the Patient Status line in white
text.
Note: x = 1, 2, 3 or 4 when used with IBP
CO
2
TECHNICAL EVENTS
CO
2
Sensor Failed The CO
2
Sensor Failed event will be posted in the Events database
if the CO
2
sensor fails and is inoperable.
The CO
2
Sensor Failed message will be shown in the Patient Status
line in white text.
CO
2
Not Calibration/
Zeroed
The CO
2
Not Calibration/Zeroed event will be posted in the
Events database if the CO
2
device is not calibrated.
The CO
2
Not Calibration/Zeroed message will be shown in the
Patient Status line in white text.
Technical Events (Continued)
MESSAGE REASON