User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 14 - 13
Alarms and Events Technical Alarms and Events
V1-V6 Lead Off The V1-6 Lead Off event will be posted in the Events database if a
V1-6 lead is not connected to the patient and a 12 lead card is in
use.
The V1-V6 Lead Off message will be shown in the Patient Status line
in white text.
RESPIRATION (ECG THORACIC IMPEDANCE) TECHNICAL EVENTS
Respiration CVA
Present
The Respiration CVA Present event will be posted in the Events
database if a CVA condition is detected for the monitored patient.
The Respiration CVA Present message will be shown in the Patient
Status line in white text.
SPO
2
(MASIMO SET® AND NELLCOR®) TECHNICAL EVENTS
SPO
2
System Check The SPO
2
System Check event will be posted in the Events
database if there is an SPO
2
system check.
The SPO
2
System Check message will be shown in the Patient Status
line in white text.
SPO
2
Communication
Error
The SPO
2
Communication Error event will be posted in the
Events database if there is an interference in the communication
between the SPO
2
and the bedside monitor.
The SPO
2
Communication Error message will be shown in the Patient
Status line in white text.
SPO
2
Failure The SPO
2
Failure event will be posted in the Events database if a
bad 5V power supply, a bad positive 15V power supply or a bad
negative 15V power supply is detected on the SPO
2
board.
The SPO
2
Failure message will be shown in the Patient Status line in
white text.
SPO
2
Board Fault The SPO
2
Board Fault event will be posted in the Events database
if the SPO
2
board fails.
The SPO
2
Board Fault message will be shown in the Patient Status
line in white text.
SPO
2
Unrecognized
Cable
The SPO
2
Unrecognized Cable event will be posted in the Events
database if the system detects an unrecognized SPO
2
cable.
The SPO
2
Unrecognized Cable message will be shown in the Patient
Status line in white text.
SPO
2
Pulse Search The SPO
2
Pulse Search event will be posted in the Events
database if the SPO
2
sensor is trying to detect a pulse.
The SPO
2
Pulse Search message will be shown in the Patient Status
line in white text.
SPO
2
Weak Pulse The SPO
2
Weak Pulse event will be posted in the Events database
if the SPO
2
sensor detects a weak pulse.
The SPO
2
Weak Pulse message will be shown in the Patient Status
line in white text.
SPO
2
(MASIMO) TECHNICAL EVENTS
SPO
2
Low Perfusion The SPO
2
Low Perfusion event will be posted in the Events
database if the SPO
2
Masimo sensor detects low perfusion.
The SPO
2
Low Perfusion message will be shown in the Patient Status
line in white text.
SPO
2
Too Much Light The SPO
2
Too Much Light event will be posted in the Events
database if the SPO
2
Masimo sensor detects too much light.
The SPO
2
Too Much Light message will be shown in the Patient Status
line in white text.
Technical Events (Continued)
MESSAGE REASON