Alarms and Events Physiological Alarms and Events
T1/T2
Units: C°
2 or 3
Factory Default: 3
High Limit Off, 35-43
o
Factory Default: Off
Off, 35-43°
Factory Default: Off
Off, 35-43°
Factory Default: Off
Low Limit Off, 26-38
o
Factory Default: Off
Off, 26-38°
Factory Default: Off
Off, 26-38°
Factory Default: Off
ST Single
Units: mm
1, 2 or 3
Factory Default: 2
High Limit N/A Off,
0.50 to10.0
Factory Default: Off
Off,
0.50 to 10.0
Factory Default: Off
Low Limit N/A Off,
(-0.5) to
(-10.0)
Factory Default: Off
Off,
(-0.5) to
(-10.0)
Factory Default: Off
ST Dual
Units: mm
1, 2 or 3
Factory Default: 2
High Limit N/A Off,
0.50-10.0
Factory Default: Off
Off,
0.50-10.0
Factory Default: Off
Low Limit N/A Off,
(-0.5) - (-10.0)
Factory Default: Off
Off,
(-0.5) - (-10.0)
Factory Default: Off
PVC
Units:
PVCs/min
2 or 3
Factory Default: 3
High Limit N/A Off, 1-30
Factory Default: Off
Off, 1-30
Factory Default: Off
Low Limit N/A N/A N/A
Arrhythmia Run 2 or 3
Factory Default: 3
High Limit N/A Off,
4-8
Factory Default: 4
Off,
4-8
Factory Default: 4
Low Limit N/A N/A N/A
Apnea Delay
Units: Seconds
2 or 3
Factory Default: 3
High Limit 10-20
Factory Default: 15
10-20
Factory Default: 15
10-60
Factory Default: 60
Low Limit
ALARM ALARM PRIORITY ALARM LIMIT NEONATE PEDIATRIC ADULT DEVICES
N/A - Not Applicable