User's Manual

Physiological Alarms and Events Alarms and Events
14 - 4 0070-00-0575-50 Panorama™ Operating Instructions
NIBP Dia.
Units: mmHg
2 or 3
Factory Default: 3
High Limit Off, 50-100
Factory Default: Off
Off, 50-100
Factory Default: Off
Off, 40-130
Factory Default: Off
Low Limit Off, 10-50
Factory Default: Off
Off, 10-50
Factory Default: Off
Off, 30-120
Factory Default: Off
NIBP Mean
Units: mmHg
2 or 3
Factory Default: 3
High Limit Off, 40-160
Factory Default: Off
Off, 50-180
Factory Default: Off
Off, 60-200
Factory Default: Off
Low Limit Off, 10-70
Factory Default: Off
Off, 10-100
Factory Default: Off
Off, 40-140
Factory Default: Off
INSP CO
2
(mmHg) 2 or 3
Factory Default: 3
High Limit Off, 5-30
Factory Default: Off
Off, 5-30
Factory Default: Off
Off, 5-30
Factory Default: Off
Low Limit N/A N/A N/A
INSP CO
2
(%) 2 or 3
Factory Default: 3
High Limit Off, 1-4
Factory Default: Off
Off, 1-4
Factory Default: Off
Off, 1-4
Factory Default: Off
Low Limit N/A N/A N/A
INSP CO
2
(kPa) 2 or 3
Factory Default: 3
High Limit Off, 1-4
Factory Default: Off
Off, 1-4
Factory Default: Off
Off, 1-4
Factory Default: Off
Low Limit N/A N/A N/A
ET CO
2
(mmHg) 2 or 3
Factory Default: 3
High Limit Off, 20-80
Factory Default: 60
Off, 20-80
Factory Default: 60
Off, 20-80
Factory Default: 60
Low Limit Off, 5-50
Factory Default: Off
Off, 5- 0
Factory Default: Off
Off, 5-50
Factory Default: Off
ET CO
2
(%) 2 or 3
Factory Default: 3
High Limit Off, 2-10
Factory Default: 5
Off, 2-10
Factory Default: 5
Off, 2-10
Factory Default: 5
Low Limit Off, 1-6
Factory Default: Off
Off, 1-6
Factory Default: Off
Off, 1-6
Factory Default: Off
ET CO
2
(kPa) 2 or 3
Factory Default: 3
High Limit Off, 2-10
Factory Default: 10
Off, 2-10
Factory Default: 10
Off, 2-10
Factory Default: 10
Low Limit Off, 1-6
Factory Default: Off
Off, 1-6
Factory Default: Off
Off, 1-6
Factory Default: Off
T1/T2
Units: F°
2 or 3
Factory Default: 3
High Limit Off, 95-110°
Factory Default: Off
Off, 95-110°
Factory Default: Off
Off, 95-110°
Factory Default: Off
Low Limit Off, 80-100°
Factory Default: Off
Off, 80-100°
Factory Default: Off
Off, 80-100°
Factory Default: Off
ALARM ALARM PRIORITY ALARM LIMIT NEONATE PEDIATRIC ADULT DEVICES
N/A - Not Applicable