User's Manual

Invalid Data Display Parameter Specifications
13 - 12 0070-00-0575-50 Panorama™ Operating Instructions
Straight Line (Waveforms)
Waveform data is replaced with a continuous straight line for the following reasons:
ECG leads are disconnected
ECG cable is not connected to the system
•SPO
2
cable is not connected to the system
Respiration parameter is disabled at the bedside monitor and the source is ECG
Respiration parameter is using a 12-lead card at the bedside monitor and the source is
ECG
•SPO
2
cable is connected and not in use
IBP cable is connected but not zeroed, or IBP is zeroing
•CO
2
filter line is not in use, or CO
2
is zeroing
•O
2
parameter is not in use
Gas module is connected but the agent is not in use
Gas module is connected but the agent is not detected