User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 13 - 9
Parameter Specifications Parameter Specifications
13.1.11 Premature Ventricular Contraction (PVC)
The Panorama Central Station supports PVC data.
NOTE: The PVC parameter displays a 0 if Arrhythmia is not in use.
13.1.12 Temp: T1 and T2
The Panorama Central Station supports two temperature measurements.
13.1.13 ST
The ST parameter provides an ST measurement using ECG leads to acquire data.
Unit of Measure: PVC/min
Resolution: 1 PVC/min
PATIENT SIZE LEADS MINIMUM VALUE MAXIMUM VALUE
Neonate 3 lead or 5 lead Not Available Not Available
12 lead Not Available Not Available
Pediatric 3 lead or 5 lead 0 PVC/min 200 PVCs/min
12 lead 0 PVC/min Less than the Current HR value
(PVCs/min)
Adult 3 lead or 5 lead 0 PVC/min 200 PVCs/min
12 lead 0 PVC/min Less than the Current HR value
(PVCs/min)
Units of Measure: F° and C°
Resolution: 0.1 F° and 0.1 C°
PATIENT SIZE MINIMUM VALUE (F°) MAXIMUM VALUE (F°)
Neonate 59 113
Pediatric 59 113
Adult 59 113
PATIENT SIZE MINIMUM VALUE (C°) MAXIMUM VALUE (C°)
Neonate 15 45
Pediatric 15 45
Adult 15 45
PARAMETER LABELS
3 lead wire set ST - I, ST - II, ST - III
5 lead wire set ST - I, ST - II, ST - III, ST - aVR, ST - aVL, ST - aVF, ST - V
12 lead card ST - I, ST - II, ST - III, ST - aVR, ST - aVL, ST - aVF, ST - V1, ST - V2, ST - V3,
ST - V4, ST - V5, ST - V6
Unit of Measure: mm
Resolution: 0.1
mm