User's Manual

Parameter Specifications Parameter Specifications
13 - 2 0070-00-0575-50 Panoramaâ„¢ Operating Instructions
13.1 Parameter Specifications
The parameters displayed on the Panorama Central Station are dependent on the type of
equipment connected to the Panorama system.
13.1.1 Heart Rate: HR
The Heart Rate parameter obtains data from ECG, SPO
2
, IBP or NIBP sources.
Heart Rate Source: ECG
The Panorama Central Station displays ECG waveforms when using 3, 5 and 12 lead
cables. The ECG leads available at the Panorama Central Station are dependent on the lead
cable connected to the patient.
Unit of Measure: Beats Per Minute (bpm)
Resolution: 1 bpm
PATIENT SIZE MINIMUM VALUE (bpm) MAXIMUM VALUE (bpm)
Neonate 30 350
Pediatric 30 300
Adult 30 300
LEAD WAVEFORM LEADS AVAILABLE
3 lead wire set I, II, III
5 lead wire set I, II, III, aVR, aVL, aVF, V
12 lead card I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Unit of Measure: mV