User's Manual

Panoramaâ„¢ Operating Instructions 0070-00-0575-50 12 - 23
Troubleshooting Messages and Issues
12.1.31 Trends Tab
This section lists some of the potential messages and issues that may occur while using the List
View of the Trends tab.
MESSAGE/ISSUE * REASON SOLUTION
There is no data in the List
view of the Trends tab.
A patient has not been
admitted to the selected tile.
Admit a patient to the selected tile or
choose a tile that already has an
admitted patient.
The Trends tab was selected
before the first interval was
reached or before the first
NIBP measurement occurred.
Wait until the first interval is reached
or until the first NIBP measurement
occurs.
A column of parameter
data is not displayed.
If the column of parameter
data exists, it has not been
scrolled into view on the
display.
Use the scroll bar located along the
bottom of the Trend list to scroll to the
desired parameter column.
The parameter is not being
monitored at the bedside.
Connect the sensor(s) for the
parameter to the patient.
No more data found
(Page Down Button)
There is no data in the Trend
database because the
Trends tab was selected
before the first interval was
reached or before the first
NIBP measurement occurred.
Wait until the first interval is reached
or until the first NIBP measurement
occurs.
The last row in the list is
already highlighted.
Select the OK button to close the
message box and return to the List
view of the Trends tab.
No more data found
(Page Up Button)
There is no data in the Trend
database because the
Trends tab was selected
before the first interval was
reached or before the first
NIBP measurement occurred.
Wait until the first interval is reached
or until the first NIBP measurement
occurs.
The first row in the list is
already highlighted.
Select the OK button to close the
message box and return to the List
view of the Trends tab.
No more events found
(Previous Event Button)
There is no event data in the
Trend database.
Wait until an event is recorded to the
Trend database.
None of the rows that are
below the currently
highlighted row contain a
parameter with an event
condition.
Select the OK button to close the
message box and return to the List
view of the Trends tab.
The last row in the list is
already highlighted.
Select the OK button to close the
message box and return to the List
view of the Trends tab.
* Messages are shown in all bold text.