User's Manual

Messages and Issues Troubleshooting
12 - 14 0070-00-0575-50 Panorama™ Operating Instructions
12.1.19 Print Setup (Patient) Tab
This section lists some of the potential messages and issues that may occur while using the
Patient Print Setup tab.
12.1.20 Print Setup (System) Tab
This section lists some of the potential messages and issues that may occur while using the
System Print Setup tab.
Please select a Print
Interval
This message appears in the
Patient Reports tab when
the Print button is pressed
before a time interval is
selected in the Full Disclosure,
Event or Trend reports.
Select the OK button to close
the dialog box. Select a print
interval in the available dialog
box.
The report will not print There are several reasons that
would cause a report not to
print. It could be a problem
with the communication
between the printer and the
Panorama Central Station or
simply a problem with the
printer.
If there is an error that appears
such as Printer out of
Paper you will need to
remedy that situation.
If the No Printer connected
or Printer not Available
error is displayed check that
the printer was properly
configured with the Panorama
Central Station. See the Printer
Configuration manual (P/N
0070-00-0561).
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.
MESSAGE/ISSUE * REASON SOLUTION
No patient Selected! This message appears in the
demographic field when either
a patient tile has not been
selected or an empty tile was
selected.
Select an active patient tile.
A waveform does not print The waveform is not selected in
the patient’s Print Setup tab.
Select the waveform in the
patient’s Print Setup tab.
* Messages are shown in all bold text.
MESSAGE/ISSUE * REASON SOLUTION
The System Setup menu
button is disabled.
The System Setup function is
already open on a Panorama
display and cannot be opened
on multiple displays.
Close the System Setup
function on the Panorama
display on which it is opened.
* Messages are shown in all bold text.