User's Manual

Messages and Issues Troubleshooting
12 - 6 0070-00-0575-50 Panorama™ Operating Instructions
12.1.9 Discharge Tab
This section lists some of the potential messages and issues that may occur while using the
Discharge tab.
12.1.10 Disclosure Tab
This section lists some of the potential messages and issues that may occur while using the
Disclosure tab.
You have reached the
maximum number of ST
licenses available
There are no more ST licenses
available
Select the OK button to close
the dialog box. ST Analysis
cannot be selected until it is
discontinued on another
patient. The System
Administrator can purchase
additional ST licenses.
Retrieve Discharged
Patient Failed: The
destination tile is
currently occupied
A patient cannot be retrieved
into an occupied tile.
Select the OK button to close
the dialog box. Select a vacant
patient tile before attempting to
retrieve a patient.
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.
MESSAGE/ISSUE * REASON SOLUTION
No patient Selected! This message appears in the
demographic field when either
no patient tile or an empty
patient tile has been selected.
Select an active patient tile.
Patient’s data and
submitted print jobs will
be deleted because at
least two of the following
fields: First Name, Last
Name or ID has not been
filled in. Do you still want
to proceed with the
discharge?
The selected patient does not
have at least two of the three
demographic fields completed.
The demographic fields include
first name, last name and ID.
• Select the Yes button to
discharge the patient.
• Select the No button to
cancel the discharge and
close the dialog box.
* Messages are shown in all bold text.
MESSAGE/ISSUE * REASON SOLUTION
No patient in the tile The selected tile is not actively
monitoring a patient.
Choose a tile that is actively
monitoring a patient.
All displays are blank and
all buttons (except
Normal Screen) are
disabled.
The selected patient tile does
not have any stored historical
data.
The selected patient tile has not been
monitoring long enough to have
stored any data. Wait for new data
to accumulate.
The selected patient tile does
not have a device attached.
Choose a tile that is actively
monitoring a patient.
All buttons (except
Normal Screen) are
disabled.
The Display Choices dialog
box is displayed.
Close the Display Choices dialog
box.
* Messages are shown in all bold text.