User's Manual

Messages and Issues Troubleshooting
12 - 4 0070-00-0575-50 Panoramaâ„¢ Operating Instructions
12.1.4 Alarm Responses Tab (Patient)
This section lists some of the potential messages and issues that may occur while using the
Patient Alarm Responses tab.
12.1.5 Bedside Tab
This section lists some of the potential messages and issues that may occur while using the
Bedside tab.
MESSAGE/ISSUE * REASON SOLUTION
No patient Selected! This message appears in the
demographic field when either
no patient tile or an empty
patient tile has been selected.
Select the OK button to close
the dialog box. Select an
active patient tile.
Communication Lost Communication between the
Panorama Central Station and
the monitoring device for the
selected patient tile has been
interrupted.
Check the connection between
the Panorama Central Station
and the monitoring device. For
example is the monitoring
device out of range?
All of the sidebar buttons
(except the Normal Screen
button) are disabled.
An empty patient tile is
selected.
Select a valid patient tile by
touching the digital or
waveform area of the patient
tile.
All Alarms Suspended The Suspend All Alarms
function is enabled. All alarm
processing for the selected tile
has been suspended.
Resume alarm processing by
pressing the Suspend All
Alarms button and disabling
the Suspend All Alarms
function.
Non-Lethal Alarms
Suspended
The Suspend Non-lethal
function is enabled. Alarm
processing for numeric,
technical and non-lethal alarms
has been suspended for the
selected tile.
Resume alarm processing by
pressing the Suspend Non-
lethal button and disabling
the Suspend Non-lethal
function.
* Messages are shown in all bold text.
MESSAGE/ISSUE * REASON SOLUTION
A waveform parameter is
not displayed.
If the waveform parameter
exists, it has not been scrolled
into view on the display.
Use the View More sidebar button
to scroll to the desired parameter
waveform.
The waveform parameter is
not being monitored at the
bedside.
Connect the sensor(s) for the
waveform parameter to the patient.
A digital parameter is not
displayed.
If the digital parameter exists,
it has not been scrolled into
view on the display.
Use the View More sidebar button
to scroll to the desired digital
parameter.
The digital parameter is not
being monitored at the
bedside.
Connect the sensor(s) for the digital
parameter to the patient.
* Messages are shown in all bold text.