User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 11 - 5
User Maintenance Telepack User Maintenance
• Use the cleaning solution to wipe and rinse the lead set.
• The cleaning solution should be wiped away with a cloth moistened with distilled water.
• A lint free cloth should be used to dry the lead set thoroughly.
CAUTION: Use of accessories and cables other than those approved by
Datascope may result in increased Emissions or decreased
Immunity of the Telepack.
NOTE: Lead sets may be immersed in the cleaning solution if they
are detached from the Telepack
Sterilization of the Lead Sets
CAUTION: Prior to sterilization, Lead sets must be removed from the
Telepack. The Telepack must not be sterilized.
Use EtO to sterilize the leadsets. The sterilization should be performed by following the
hospital-approved procedure for EtO sterilization, such as those recommended by AAMI. The
leadsets remain effective up to 10 exposures to EtO sterilization cycles.
NOTE: Damage to the lead sets will result if steam or gamma
radiation is used for the sterilization process.
11.2.3 Disposal of the Telepack Device
Adhere to the following guidelines when disposing of any Telepack device.
• The Telepack should be disposed of in a manner adhering to the state laws and federal
regulations regarding disposal of electronic and computer accessories. For example, the
State of Minnesota (US) prohibits disposing of electronic and computer accessories in the
trash. Follow appropriate state/federal laws.
• If your state/country laws recommend recycling, you can dispose of the device by
following the recycling program offered by state/federal run agencies.
11.2.4 Disposing of Batteries
Adhere to the following note when disposing the batteries.
NOTE: The alkaline batteries used in the Telepack may be subject to
local regulations regarding disposal. At the end of the
battery life, never dispose of the batteries in a fire but
dispose of the batteries in accordance with any local
regulations.