User's Manual

Telepack User Maintenance User Maintenance
11 - 4 0070-00-0575-50 Panorama™ Operating Instructions
The cleaning solutions that can be used are:
Soap and water
Glutaraldehyde 2% (for example, Cidex)
Dilute chlorine bleach (sodium hypochlorite), 10% solution, freshly prepared in past 24
hours
Isopropyl alcohol 70%
•Ethyl alcohol
Guidelines for Cleaning the Telepack
Remove the batteries, and reattach the end cap of the battery compartment firmly. Clean
the empty Telepack using a cloth moistened with cleaning solution, by gently wiping the
outside of the Telepack.
Remove the ECG lead wire from the device to clean around the ECG connector.
NOTE: Do not immerse the Telepack in the cleaning solution.
Any excess cleaning solution should be wiped away with a cloth moistened with distilled
water.
A lint free cloth should be used to dry the Telepack unit thoroughly.
11.2.2.1 Cleaning the Battery Compartment
To clean the battery compartment, follow these guidelines:
Remove the batteries from the battery compartment by twisting the endcap in a counter-
clockwise direction. Visually inspect for fluids or foreign matter.
Use a gauze pad or long handle swab moistened with soap, water or alcohol to clean
the battery compartment.
Excess cleaning solution should be wiped away with a cloth moistened with distilled
water.
A lint free cloth should be used to wipe the inside of the battery compartment. Air dry the
battery compartment prior to re-inserting the batteries.
NOTE: Ensure that the battery contact leads and the battery
compartment are dry prior to re-inserting the batteries for
use.
11.2.2.2 Cleaning the ECG Lead Sets
To clean the ECG lead sets, use any one of the following cleaning solutions:
Soap and water
Glutaraldehyde 2% (for example, Cidex)
Dilute chlorine bleach (sodium hypochlorite), 10% solution, freshly made in past 24 hours
Ethyl Alcohol 30% (for example, green soap tincture)
To clean the ECG lead sets, remove them from the Telepack, and follow these guidelines: