User's Manual

Troubleshooting Telepack (Optional)
10 - 12 0070-00-0575-50 Panorama™ Operating Instructions
10.6 Troubleshooting
This section lists some of the potential messages and issues that may occur while using a
Telepack device.
MESSAGE/ISSUE * REASON SOLUTION
Telepack/LED’s not turning on Batteries are not installed
correctly, batteries are
depleted, or batteries
corroded.
Visually inspect the battery
compartment for any foreign
object.
Replace the batteries.
Telepack not transmitting data
to the Panorama Central
Station.
Batteries are discharged Press the Test/CHECK button
to verify that the batteries are
depleted. Replace as
necessary.
Flashing Battery Status LED
indicator light
Batteries in the Telepack are
weakening
Replace batteries
Error: Could not receive
data message, while
attempting to program the
Telepack
Telepack is not connected
properly, or the Panorama
Central Station is unable to
detect the Telepack.
Verify that the service mode
cable is properly connected to
the device.
Retry programming.
Check batteries.
Unlit ECG Lead LED indicator
light
Telepack has no lead
connectivity.
Connect the electrode.
Check batteries.
Communication Lost Telepack’s radio signal is out of
range with the wireless
network
Verify the Telepack is within
100 feet of a network access
point.
Check batteries.
Flashing Link Status LED
indicator light
Telepack is not communicating
with the Panorama Central
Station
Verify that Device ID is
registered on the Panorama
Central Station.
Verify that Panorama Central
Station is functioning.
* Messages are shown in all bold text.