User's Manual

Wireless Tab System Setup Functions
9 - 48 0070-00-0575-50 Panorama™ Operating Instructions
Data is shown in the Telepack IP Address field when the Wireless tab’s Read sidebar
button is pressed and has a factory default setting of blank. If the read was not
successful the Telepack IP Address field will remain blank. To view the connection
status for the wireless device check the message in the Connect field.
A Telepack IP Address can be programmed in a connected device by selecting the
Program sidebar button.
NOTE: The Telepack IP Address field will not be shown when the
device cannot be recognized or there is a Read error.
Subnet Mask IP Address
The Subnet Mask IP Address field displays the Subnet Mask IP address for the assigned
wireless device. It sets the subnet mask of the attached device.
Data is shown in the Subnet Mask IP Address field when the Wireless tab’s Read
sidebar button is pressed and has a factory default setting of blank. If the read was not
successful the Subnet Mask IP Address field will remain blank. To view the connection
status of the wireless device check the message in the Connect field.
A Subnet Mask IP Address can be programmed in a connected device by selecting the
Program sidebar button.
NOTE: The Subnet Mask IP Address field will not be shown when
the device cannot be recognized or there is a Read error.
Network Name
The Network Name field displays the name of the network on which the assigned device
resides. It sets the Network Name of the attached device.
Data is shown in the Network Name field when the Wireless tab’s Read sidebar
button is pressed and has a factory default setting of blank. If the read was not
successful the Network Name field will remain blank. To view the connection status
check the message in the Connect field.
A Network Name can be assigned to a connected device by selecting the Program
sidebar button.
NOTE: The Network Name field will not be shown when the device
cannot be recognized or there is a Read error.
Lead Selection
The Lead Selection button displays the ECG Lead to be monitored by the assigned device.
It sets the lead to be monitored by the attached device. ECG lead selection options include
Lead I, Lead II and Lead III.
Data is shown in the Lead Selection field when the Wireless tab’s Read sidebar button
is selected. Lead Selection has a factory default setting of blank. If the read was not
successful the Lead Selection field will remain blank. To view the connection status
check the message in the Connect field.
The Lead to be monitored by a connected device can be assigned by selecting the
Program sidebar button.
NOTE: The Lead Selection button will not be shown when the
device cannot be recognized or there is a Read error.