User's Manual

Wave Gain Tab (System) System Setup Functions
9 - 44 0070-00-0575-50 Panorama™ Operating Instructions
9.15 Wave Gain Tab (System)
The Wave Gain tab (FIGURE 9-19) sets the system default wave gain settings for the
Panorama Central Station.
For information on the setting of patient wave gains, refer to the ‘‘Wave Gain Tab (Patient)’’
on page 5-37.
NOTE: The changes made in the Wave Gain tab (System) will apply
to newly admitted patients.
This section includes the following:
• Accessing the Wave Gain Tab
• Adjusting the System Default Wave Gain Settings
• Sidebar Button - Normal Screen
• Troubleshooting
9.15.1 Accessing the Wave Gain Tab
1. From the menu bar, select the System Setup button. The first set of System Setup tabs
is displayed.
2. Select the Installation Setup tab. The Installation Setup tab is displayed.
3. Enter the correct System password. The next set of System Setup tabs is displayed.
4. Select the More tab. The next set of System Setup tabs is displayed.
5. Select the Wave Gain tab.
FIGURE 9-19 Wave Gain Tab (System)
9.15.2 Adjusting the System Default Wave Gain Settings
This section outlines entering and modifying the system default wave gain settings.
1. Select the ECG mm/mV button until the desired ECG wave gain is displayed.
The settings available for the ECG wave gain include 2.5, 5.0, 10.0, 20.0 and 30.0
mm/mV. The factory default setting is 10.0 mm/mV.