User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 9 - 41
System Setup Functions Unit Choices Tab
7. Select the Demographic Line Position 4 button to select the system default
demographic data to be shown in the fourth demographic line position of the Patient
Status line in the main screen of the Panorama Central Station.
Demographic Line Position 4 options include ID, Demo Off, Last Name, First Name
and Bed. The factory default setting is ID.
NOTE: Changing the Demographic Line Position 4 option will affect
all patients with the next update of the Patient Status line in
the patient tile.
8. Select the Height Units button to select the system default unit of measurement to be
used for the height variable in the Panorama Central Station. Height Unit options
include cm (centimeters) and in (inches). The factory default setting is cm.
NOTE: Changing the Height Units option will affect new Telepack
patients upon entering the Demographics tab.
9. Select the Weight Units button to select the system default unit of measurement to be
used for measuring weight at the Panorama Central Station. Weight Unit options include
lbs (pounds) and kg (kilograms). The factory default setting is lbs.
NOTE: Changing the Weight Units option will affect new Telepack
patients upon entering the Demographics tab.
10. Select the Temperature Units button to select the system default temperature unit to
be used for measuring temperature at the Panorama Central Station. Temperature unit
options include F (Fahrenheit) and C (Celsius). The factory default setting is F.
NOTE: Changing the Temperature Units option will affect all new
patients upon entering the System Alarm Limits tab.
11. Select the CO2 button to select the system default CO
2
unit to be used for measuring
CO
2
at the Panorama Central Station. CO
2
unit options include %, kPa and mmHg. The
factory default setting is %.
NOTE: Changing the CO2 Units option will affect all new patients
upon entering either the System Alarm Limits or System
Wave Gain tabs.
9.14.3 Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this tab.
Arrhythmia (Telepack)
The Arrhythmia sidebar button enables and disables Arrhythmia Analysis as a system default
for patient tiles that have a Telepack as the assigned device. The availability of the
Arrhythmia function is based on the availability of Arrhythmia licenses. Contact your System
Administrator for additional information on Arrhythmia licenses. To enable Arrhythmia for a
Telepack device, follow this step:
Select the Arrhythmia sidebar button in the Unit Choices tab.
A highlighted white button with blue text indicates that Arrhythmia Analysis is enabled.
An unhighlighted navy blue button with white text indicates that Arrhythmia Analysis is
disabled.