User's Manual

Unit Choices Tab System Setup Functions
9 - 40 0070-00-0575-50 Panoramaâ„¢ Operating Instructions
NOTE: Changing the Pacer Enhancement option will affect all
newly admitted patients and will have no affect on patients
currently being monitored.
2. Select the Pacer Reject button to select the system default pacer reject period. Pacer
Reject adjusts the time period at which pacemaker-induced ECG complexes are filtered
for ECG sampling processes.
Pacer Reject options include 0 to 80 milliseconds (ms) in five-second increments. The
factory default setting is 40 ms.
NOTE: Changing the Pacer Reject option will affect all newly
admitted patients and will have no affect on patients
currently being monitored.
3. Select the Trend Interval button to select the system default interval period for trend
storage. The trend interval period is the amount of time that has elapsed between trend
collections at the Panorama Central Station.
Trend Interval periods include Off, 1 minute, 5 minutes, 10 minutes, 15 minutes, 30
minutes, 1 hour and 2 hours. The factory default setting is 1 minute.
4. Select the Demographic Line Position 1 button to select the system default
demographic data to be shown in the first demographic line position of the Patient
Status line in the main screen of the Panorama Central Station.
Demographic Line Position 1 options include Last Name, First Name, Bed, ID and
Demo Off. The factory default setting is Last Name.
NOTE: Changing the Demographic Line Position 1 option will affect
all patients with the next update of the Patient Status line in
the patient tile.
5. Select the Demographic Line Position 2 button to select the system default
demographic data to be shown in the second demographic line position of the Patient
Status line in the main screen of the Panorama Central Station.
Demographic Line Position 2 options include First Name, Bed, ID, Demo Off and
Last Name. The factory default setting is First Name.
NOTE: Changing the Demographic Line Position 2 option will affect
all patients with the next update of the Patient Status line in
the patient tile.
6. Select the Demographic Line Position 3 button to select the system default
demographic data to be shown in the third demographic line position of the Patient
Status line in the main screen of the Panorama Central Station.
Demographic Line Position 3 options include Bed, ID, Demo Off, Last Name and
First Name. The factory default setting is Bed.
NOTE: Changing the Demographic Line Position 3 option will affect
all patients with the next update of the Patient Status line in
the patient tile.