User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 9 - 39
System Setup Functions Unit Choices Tab
9.14 Unit Choices Tab
The Unit Choices tab (FIGURE 9-17) configures pacemaker settings, the order of
demographic data and to select units of measure for the Panorama Central Station.
This section includes the following:
Accessing the Unit Choices Tab
Setting the Unit Choices
Sidebar Buttons
Troubleshooting
9.14.1 Accessing the Unit Choices Tab
1. From the menu bar, select the System Setup button. The first set of System Setup tabs
is displayed.
2. Select the Installation Setup tab. The Installation Setup tab is displayed.
3. Enter the correct System password. The next set of System Setup tabs is displayed.
4. Select the More tab. The next set of System Setup tabs is displayed.
5. Select the Unit Choices tab.
FIGURE 9-17 Unit Choices Tab
9.14.2 Setting the Unit Choices
This section outlines selecting system defaults for pacemaker settings, order of demographic
data and units of measurement.
1. Select the Pacer Enhancement button to set the system default setting for pacer
enhancement. Pacer Enhancement will visually mark pacemaker-induced ECG
complexes with a colored vertical spike.
The Pacer Enhancement settings for a specific patient can be changed in the Pacer
Settings dialog box in the Demographics tab. Pacer Enhancement options include ON
and OFF. The factory default setting is ON.
•The ON option indicates that Pacer Enhancement is enabled.
•The OFF option indicates that Pacer Enhancement is disabled.