User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 9 - 25
System Setup Functions Equipment Setup Tab
Illegal Device Label.
Device Label must be
defined. Please try again.
This message appears when
the Done button is pressed in
the Equipment Setup tab
before a device label is
selected.
Select the OK button to close
the dialog box. Select a device
label in the Device Label
dialog box.
Device Label already
exists! Please enter a
new one
This message appears when a
duplicate device label is
entered in the Device Label
field.
Select the OK button to close
the dialog box. Enter a new
device label in the Device
Label field.
Invalid characters
entered. Please enter a
new label.
This message appears when
invalid characters have been
entered in the Device Label.
The device label can only
contain alphanumeric
characters and underscores
(_).
Select the OK button to close
the dialog box. Enter a valid
Device label.
Illegal Device ID. Please
enter a new one.
This message appears when
the Done button is pressed in
the Equipment Setup tab
before a device ID is selected.
Select the OK button to close
the dialog box. Enter a Device
ID in the Device ID dialog box.
Device ID must be 8
characters in length.
This message appears when
the Device ID that was entered
is less than 8 characters in
length. Please refer to the
Device ID setup rules in the
Setting up New Equipment
section.
Select the OK button to close
the dialog box. Enter a Device
ID in the Device ID dialog box.
Device ID must start with
the proper device prefix
This message appears when
the incorrect device prefix is
entered in the Device ID box.
Select the OK button to close
the dialog box. Enter the
correct Device ID prefix.
Suffix of Device ID last
character must be ‘x’ or
‘X’.
This message appears when
the incorrect device suffix is
entered in the Device ID box.
Select the OK button to close
the dialog box. Enter the
correct Device ID suffix.
This Device ID already
exists!
This message appears when a
duplicate Device ID is entered
in the Device ID field.
Select the OK button to close
the dialog box. Enter a new
Device ID in the Device ID
field.
You have reached the
maximum number of
channel licenses
available.
This message appears when all
available channels on the
Panorama Central Station are
occupied.
Select the OK button to close
the dialog box. Delete a piece
of equipment from an existing
channel if it is no longer in use.
The system administrator can
also purchase additional
channel licenses.
This version of software
does not support more
than 16 monitored
patients.
This message appears when
the maximum number of
patients is being monitored on
the Panorama Central Station.
Select the OK button to close
the dialog box. Discharge a
patient if the patient is no
longer being monitored.
System Setup menu button is
disabled.
The System Setup function is
already open on a Panorama
display and cannot be opened
on multiple displays.
Close the System Setup
function on the Panorama
display on which it is opened.
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.