User's Manual

System Alarms Tab System Setup Functions
9 - 12 0070-00-0575-50 Panorama™ Operating Instructions
NOTE: Changes to the Latch Alarms option will only affect new
alarm violations.
2. Select the Show HR Limits in Tile toggle button to either enable or disable the
display of heart rate alarm limits in the Bedside tab. Show HR Limits in Tile options
include ON and OFF. The factory default setting is ON. The Heart Rate alarm limits
define the range in which heart rate values are measured.
•The ON option indicates that HR limits will be displayed in the Bedside tab.
•The OFF options indicates that HR limits will not be displayed in the Bedside tab.
NOTE: Changes to the Show HR Limits in Tile option will affect all
patients.
3. Select the Mute Duration button to select the mute duration period for alarms in the
Panorama Central Station. The mute duration period is the amount of time for which
alarming parameters are silenced. Mute Duration periods include 10, 15, 30, 45, 60
and 120 seconds. The factory default setting is 30 seconds.
NOTE: Changes to the Mute Duration period will take effect within
ten seconds of the change.
4. Select the Technical Event Sounds toggle button to enable or disable sound during a
technical event. Technical Events Sounds options include ON and OFF. The factory
default setting is ON.
•The ON option indicates that Technical Events Sounds will be heard during a
technical event.
•The OFF options indicates that Technical Events Sounds will not be heard during
a technical event.
NOTE: Changes to the Technical Events Sounds option will affect all
patients within 10 seconds of the change.
5. Select the Apnea Latching toggle button to either enable or disable the apnea
latching function. Apnea latching is the ability to set Apnea alarms to continue the alarm
behaviors until the alarm is acknowledged by the user. Apnea latching options include
ON and OFF. The factory default setting is ON.
•The ON option indicates that Apnea Latching will be enabled.
•The OFF options indicates that Apnea Latching will not be enabled.
NOTE: Changes to the Apnea option will affect all new Apnea
alarms.
6. Select the Password Protection toggle button to either enable or disable the
password protection for the Patient Alarms tab in the Panorama Central Station.
Password Protection options include ON and OFF. The factory default setting is
ON.
•The ON option indicates that Password Protection will be enabled in the Patient
Alarms tab.
•The OFF options indicates that Password Protection will not be enabled in the
Patient Alarms tab.
NOTE: Changes to the Password Protection option will affect all
patients upon entering the Patient Alarms tab.