Additional Reports Report Functions
8 - 48 0070-00-0575-50 Panorama™ Operating Instructions
• Unconfirmed Report: The 12-lead Report is marked as an Unconfirmed Report when
interpretation strings are printed on the report.
8.4.5.2 Waveform Data
The 12-lead Report section displays 2.5 seconds of waveform data for each of the 12 leads
monitored at the bedside monitor, a 10-second ECG rhythm strip for Lead II, a calibration
pulse and a wave gain setting of 1 millivolt.
8.4.5.3 Printing a 12-lead Report
The 12-lead Report prints when a print request is received from a connected bedside monitor.
This report prints in landscape orientation.
Before this report is requested, the following must be configured at the bedside monitor:
• 12-lead card must be in use
• The monitor must be in the View All ECG Mode
• Remote printing must be enabled
To print the report, select the STRIP button from the bedside monitor. A print request is sent
from the bedside monitor to the Panorama Central Station’s selected printer.
NOTE: Requesting this report posts a print job in the system Print
Status tab. For more information, refer to the “Print Status
Tab” on page 8-33.
8.4.6 Telepack Error Log Report
The Telepack Error Log Report is a one page report that shows all logged errors for any
wireless device connected to the Panorama Central Station in service mode. This report
includes a header, error log data and a footer.