User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 8 - 45
Report Functions Additional Reports
8.4.4.2 Digital Data
The Trend Graphic Report’s digital data section displays measurements for the digital data
parameters currently available for the selected patient tile. The displayed parameter order is
determined by the system Unit Priorities tab. The digital data section contains the
following:
A parameter label
A numeric measurement
A unit of measure label
A heart rate (HR) value
Dashes (--) to indicate missing or invalid digital data.
Square brackets [XX] to indicate alarming parameter values
NOTE: Digital Data is only shown on the first page of the report.
8.4.4.3 Graphic Data
The graphic data section of the Trend Graphic Report displays a cursor time, parameter list
and a graphical representation of the parameter measurements that occurred during the
selected time period.
The cursor time represents the time that was selected in the Trends (Graphic) tab. The
cursor time in the report follows the system Date/Time format.
The parameters shown in this report display on the left side of the report. Each parameter
has its own graphical layout which includes a parameter label, a unit of measure label, a
scale bar and the scale bar range of values.
For historical purposes, the Graphic Data section of the report shows all the selected
parameters, even if the parameters are currently turned off.
The report shows a vertical line on the graph to represent the occurrence of a
physiological alarm.
The report shows nine time periods across the top of the graph for a more representative
look at the parameter data. The time periods will be shown in the system default time/
date format.
The time periods shown in the report are equal sections of the total time interval selected
in the graphic report. For example, if a 4 hour interval is selected, graphical data is
plotted in 30 minute increments and if an 8 hour interval is selected, data is plotted in 60
minute increments. This report is capable of plotting graphical data in 2, 4, 6 and 8 hour
time intervals. The report is printed in the time interval specified in the Trends (Graphic)
tab.
Time intervals are changed in the Trends (Graphic) tab by selecting the Zoom In and
Zoom Out buttons.