User's Manual

Additional Reports Report Functions
8 - 42 0070-00-0575-50 Panorama™ Operating Instructions
ECG Waveforms
The ECG waveforms shown in the Full Disclosure Zoom In Report are limited to the ECG lead
wire set currently in use and on the waveforms selected in the Print Setup tab. The ECG
waveform report displays with the following:
An ECG lead label
A wave gain scale bar and a calibration pulse for wave gains less than 30 mm/mV
A calibrated grid
Current ECG monitoring label
A continuous straight line for active ECG leads that are not connected
Non-ECG Waveforms
The non-ECG waveforms shown in the Full Disclosure Zoom In Report are based on the non-
ECG waveforms that are in use and the non-ECG waveforms that are selected to be printed.
All non-ECG waveforms in the report contain a parameter label and a unit of measure label.
Non-ECG waveforms in the report may contain a wave gain scale bar. If a non-ECG
waveform is unavailable at the time of the print request, the report prints blank data for the
specified time period.
If the IBP waveform was labeled at the bedside monitor, the report contains that IBP label.
If the Anesthetic Agent was identified at the bedside monitor, the report contains that
Agent label.
8.4.2.4 Footer
The Full Disclosure Zoom In Report contains a footer at the bottom of every page. The footer
includes the following:
The date and time of the print request. The format displayed is taken from the System
Date/Time tab.
A centered page number.
The name of the Panorama Central Station where the report was generated.
8.4.3 Printing a Full Disclosure Zoom In Report
The Full Disclosure Zoom In Report is printed in landscape orientation. Follow these steps to
print this report:
1. From the Disclosure tab, use the Zoom In button to magnify a portion of the
compressed waveform view.
2. Select the Print button.
NOTE: Requesting this report posts a print job in the system Print
Status tab. For more information, refer to the “Print Status
Tab” on page 8-33.