User's Manual

Panorama™ Operating Instructions 0070-00-0575-50 8 - 25
Report Functions Patient Reports Tab
The third column of this report displays the High alarm limit for the parameter specified.
The fourth column of this report displays the Low alarm limit for the parameter specified.
The fifth column of this report displays the alarm priority assigned to the parameter
specified.
The sixth column of this report displays the Print on Alarm response assigned to the
parameter specified.
The seventh column of this report displays the Save to Event alarm response that has been
assigned to the parameter specified.
8.1.7.3 Footer
The Patient Alarm Report’s contains a footer at the bottom of the page. The footer includes the
following:
The date and time of the print request. The format displayed is taken from the System
Date/Time tab.
A centered page number.
The name of the Panorama Central Station where the report was generated.
8.1.7.4 Printing a Patient Alarm Report
This Patient Alarm Report is printed in portrait orientation, using any of the following
methods:
Select the Patient Alarm Report button in the Patient Reports tab.
Select the Print sidebar button in the patient’s Alarm Limits tab.
NOTE: Requesting this report posts a print job in the system Print
Status tab.
8.1.8 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Patient Reports tab.
MESSAGE/ISSUE * REASON SOLUTION
All report buttons are disabled. A patient is not currently
selected.
Select a patient before trying to
print a report.
The ST REPORT button is
disabled.
The selected patient does not
have ST enabled.
Enable ST for the selected
patient before trying to print an
ST report.
* Messages are shown in all bold text.