Panorama™ Operating Instructions 0070-00-0575-50 8 - 9
Report Functions Patient Reports Tab
Page headers on subsequent pages of the report provide the report title, the date and time
the data was collected, and limited demographic information about the patient (first and last
name, identification number and bed number), if it is available.
If the demographic data was not previously entered in the system, only the demographic
label is shown.
8.1.3.2 Waveform Data
The ECG waveforms shown in the Full Disclosure Report are based on the ECG lead wire set
currently in use and on the waveforms selected in the Print Setup tab.
The first page of this report contains up to two waveforms while subsequent pages contain up
to three waveforms. Each waveform in this report represents approximately 10 seconds of
data; five seconds prior to the time of the print request and 5 seconds after the print request.
A waveform label (i.e., ECG Lead II) is shown above each strip to identify the data source
and the unit of measurement (i.e.,%, RPM, BPM, mmHg, kPa). Pleth, ECG and Respiration do
not display a unit of measurement. The waveforms displayed in the report are shown on a
calibrated grid and contain a scale bar.
For pacemaker patients that have Pacer Enhancement enabled, the report displays a vertical
bar to indicate the firing of the pacemaker impulse.
ECG Waveform Format
The ECG waveform format for the Full Disclosure Report includes five columns. This format is
used for reports that show ECG parameter waveforms.
• The first column contains the time (hour and minute) of the disclosure record. The time
format follows the system format specified in the Date/Time tab.
• The second column contains a heart rate (HR) value. The heart rate value shown is from
the last second of data that was collected during the disclosure record. If disclosure data
is not available for the time period specified, dashes (---)are shown in the place of a heart
rate value.
• The third column contains a PVC/min. value. The PVC value shown is the value that was
obtained during the disclosure record. If disclosure data is not available for the time
period specified, dashes are shown in the place of a PVC value.
• The fourth column contains a label for the waveform data being displayed in the report.
The waveform label shown is the source of data for the disclosure record.
• The fifth column contains the events that occurred during the period of the disclosure
record. If multiple events occur during the time period specified, the event with the highest
priority is shown. If no events occur during the time period specified, this column remains
empty.
For information regarding the type of events that occur in the Full Disclosure Report,
“Physiological Alarms and Events” on page 14-2.
Non-ECG Waveform Format
The non-ECG waveform format for the Full Disclosure Report includes four columns. This
format is used for reports that show Non-ECG parameter waveforms.